Rhopoint Appearance Elements Help Help

7. Maintenance and Support

Cleaning Routines

To ensure accurate measurements and optimal performance of your Rhopoint glossmeter, follow these cleaning procedures:

  1. Instrument Optics:

    • Inspect the optics before each calibration.

    • Remove dust or debris using dry, clean air.

    • Never touch the optics without wearing suitable cotton gloves.

    • If permanent marks or scratches are present on the lenses, return the instrument to an authorized Rhopoint service center[^3].

  2. Calibration Tile:

    • Inspect the calibration standards before each use.

    • Remove fingerprints and dust with the supplied optic cleaning cloth.

    • If the tile is scratched or damaged, it is not suitable for calibration and must be replaced[^3].

Annual Calibration Recommendations

Rhopoint Instruments recommends annual calibration of both the glossmeter and its reference tile:

  • Send your glossmeter and tile to a Rhopoint Approved Service Agent yearly.

  • Annual calibration ensures accuracy and compliance with current industry standards.

  • Calibration maintains optimal performance and helps meet desired aesthetic and quality standards.

Calibration Options:

  • ISO 17025 UKAS certified calibration for Rhopoint Instruments glossmeters.

  • Traceable calibration for all other glossmeter brands.

Support Contact Details

United Kingdom (Head Office)

Rhopoint Instruments Ltd.

  • Address: Rhopoint House, Enviro 21 Park, Queensway Avenue South, St. Leonards-on-Sea, East Sussex, TN38 9AG, UK

  • Hours: Monday to Thursday: 08:30 to 17:00 GMT, Friday: 07:30 to 13:00 GMT

  • Email: sales@rhopointinstruments.com, support@rhopointinstruments.com

United States

Rhopoint Americas Inc.

  • Address: 1000 John R Road, Suite 209, Troy, MI, 48083, United States

  • Hours: Monday to Friday: 09:00 to 17:00 (Local time)

Germany

Rhopoint Instruments GmbH

  • Address: Am Weiglfeld 28, 83629 Weyarn, Germany

  • Hours: Monday to Thursday: 08:00 to 17:00, Friday: 08:00 to 14:00 (Local time)

Additional Support

  • Live chat support is available during UK office hours.

  • For assistance outside office hours, email support@rhopointinstruments.com.

  • Response time for email inquiries is typically within 24 hours.

Remember to regularly maintain and calibrate your Rhopoint glossmeter to ensure accurate and reliable measurements for your quality control processes.

Last modified: 07 March 2025